I nstruments 14 Round Pin 9304-3005 B. Proximal Tibial Resection Place the knee in 90 degrees flexion. Secure around the ankle joint with the clamps on the distal portion of the Tibial EM Alignment Guide proximal to the malleoli. Rotate the height locking knob to unlock position, and adjust the Tibial EM Alignment Guide to the approximate length of the tibia. Tibial Resection Guide Left: 9404-2001 Right: 9404-2002 Tibial EM Alignment Guide 9404-2101 Align the central groove on the Tibial Resection Guide with the medial one third of the tibial tubercle. Drill a pilot hole with the 3.2 mm Drill through the vertical slot, and secure the proximal rotation with a Round Pin . Round Pin 3.2 mm Drill 9304-3100