I nstruments 21 Fix with Threaded Pins D. Femoral Sizing and Chamfer Resection Placement of the AiO Block Place the AiO Block against the cut distal surface of the femur with the posterior condyles seated on the posterior feet. Then secure the block with two Threaded Pins through the holes that connects to the posterior feet. AiO Block 9304-7101 Threaded Pin 30 mm/50 mm 9304-3003/9304-3004 Gap Spacer 1 mm/2 mm 9304-5301/9304-5302 Optional technique for excessive worn posterior condyles: � If one posterior condyle is worn excessively, use the 1 mm or 2 mm Gap Spacer to restore desired posterior condylar thickness.