United Orthopedic Corporation

Straight Stem Drill 9403-3214-RA (CMA Tray) Tibial Drill Guide, CMA 9403-2105-RB (CMA Tray) I nstruments Tibial Punch Handle 9404-1101 To ensure the stability of tibial component, an extended 30 mm distal stem for the implant is recommended. Advance the Straight Stem Drill through the CMA Tibial Drill Guide , until the depth reaches the laser mark of the “0-4” or “5-7” line according to the selected size of the baseplate trial. The drill and drill guide are then removed. Continue with tibia punching to finish the fin slot preparation. (Referring to page. 39) Appendix : Size 5- 7 Size 0- 4 Tibial Punch #0~#2: 9404-6001 #3~#4: 9404-6002 #5~#7: 9404-6003 Tibial Punch Guide 9404-2102 56