The extension and flexion joint gaps may be evaluated at this time with the Gap Gauge 24 . The 9 mm Gap Gauge 24 is initially selected to assess both the extension and flexion joint gaps. If a thicker gap is required, combine additional Gap Gauge 24 blocks with different thicknesses and test again. The range of thickness is from 9 mm to 18 mm. If neither the extension and flexion gaps nor soft tissue tension shows any problem, insert the femoral and tibial trial to test the knee mobility and their relative positions. NOTE : The Alignment Rod 3 may be inserted through the Gap Gauge 24 handle to assess the extramedullary alignment in both extension and flexion. B. Tibial Preparation B.3. Extension and Flexion Gaps Confirmation 11 24 Gap Gauge 9403-7009 9403-7011 9403-7013 9403-7015 9403-7018