Acetabular Preparation Acetabular reaming Appropriate reaming of the acetabulum is important for the cup to be fully seated within. All articular cartilage, osteophytes, and any soft tissues should be removed throughout the reaming process. It is important to understand the labeled size on the U-Motion II + acetabular instrument (Reamer and Cup Trial) is an absolute dimension, however the implant has 0.35 mm surface coating thickness in each side. For example, a 58 mm U-Motion II + cup measures 58.7 mm at the rim (Fig. 2), thus under-reaming of the cavity by 1 mm or line-to-line reaming should depend on bone quality. Hold the Cup Reamer Handle 1 at abduction of 40°- 45° and anteversion of 15°- 20° (Fig. 3). Utilize the smallest Cup Reamer 2 to begin acetabular reaming, then gradually proceed with enlarged reamers in 1 mm increments until the anticipated size is achieved. 1 1. Cup Reamer Handle, EZ Clean 9206 - 1101 2. Cup Reamer C/N varies by sizes Fig. 3 Fig. 2 Acetabular Reamer 58 mm Reamer has a 58 mm O.D. Cup Trial 58 mm Trial has a 58 mm O.D. Acetabular Cup 58 mm Cup has a 58.7 mm O.D.