24 To impact the Tibial Baseplate, it is recommended to carefully introduce and align the stem of the implant into the prepared stem hole. The implant may be positioned by hand or by using the Tibial Baseplate Driver 59 . Once the Tibial Baseplate is advanced sufficiently, the Tibal Baseplate Impactor 60 may then be used to complete seating of the implant. To impact the Femoral Implant, the Femoral Driver 26 is assembled onto the Femoral Implant. Carefully align the femoral implant with the distal femur to assure correct advancement and seating of the implant. The Femoral Impactor 27 may also be used for seating if desired. The Patellar Implant is firstly seated by hand, carefully aligning the implant peg(s) with the prepared bone bed. The Patellar Resection Clamp 47 is equipped with the Patella Cement Clamp Adaptor 61 . This assembly is then used to fully seat the cemented implant in a standard fashion. Femoral Tibial trialing may now be performed again if desired. F. Implant Fixation F.2. Implant Fixation 59 Tibial Baseplate Driver 9403-5101-RC 60 Tibial Baseplate Impactor 9403-5102-RF 61 Patella Cement Clamp Adaptor 9401-5312-RD -Caution: Femoral Driver is designed to position the implant and trial, not for final impaction. Please impact gently to avoid instrument breakage.